02/24/99 - by D. Pease - v1
We summarize on the following pages the current Analysis of the HRC Point Spread Function from the XRCF data. The PSF data set consists of all specified PSF tests, Effective Area measurements at off-axis angles greater than 8.0 arcmin (performed in focus), and the Count Rate Linearity tests for comparison to non-dithered data. We have calculated the FWHM and Encircled Energy of the PSF. Some details can be found in the HRC PSF Analysis Software section.
The data were processed with Version 1 of the Degap Parameters.
EIPS | HIREFS | DCM |
---|---|---|
0.277 keV (C-Ka) | 0.392 keV | 2.560 keV |
0.784 keV | 3.500 keV | |
0.905 keV | 4.960 keV | |
1.176 keV | 7.000 keV | |
1.487 keV | ||
1.810 keV |
See also the IPI Team's HRC-I PSF analysis.
EIPS | HIREFS | DCM |
---|---|---|
0.277 keV (C-Ka) | 0.392 keV | 2.560 keV |
0.784 keV | 3.500 keV | |
0.905 keV | 4.960 keV | |
1.176 keV | 7.000 keV | |
1.487 keV | ||
1.810 keV |
See also the IPI Team's HRC-S PSF analysis.