The default value of the commanded event-width threshold was changed for both
the HRC-I and HRC-S from 3 to 2 at 2000:279:12:57. This change occurred
between two HRC observations - ObsIDs 1857 and 1853.
The change was made so that two of the bits in the event status word could be
used to help unambiguously identify those events that require a correction to
the telemetered crossed-grid charge detector signals in order to reduce image
blurring. A future version of the CIAO tool hrc_process_events will use these
bits. The original event-width threshold was chosen to flag events that were
likely to be due to charged particles but other screening tests performed on
the events provide this flagging.
Effective with the instrument configuration change, a change was made to the
standard level 2 processing pipeline. The event status bit mask used to filter
when making the level 2 event-list was modified so ignore the state of the two
relevant bits in the event-status word; this change was effective in software
version R4CU5UPD11.1. Users should be cautioned not to use the bit-mask
defined in earlier versions of the software on HRC data taken after
2000:279:12:57.
Last modified:
09/25/12
10/05/00 - HRC Command Event Width Threshhold