Step	Title                    	Time	Command     	Command     	Cmd         	Seq         	Telemetry   	Telemetry   	Expected    	Units 	Telemetry   	Other       	Crit        	Description              	Notes       	RT 	Tlm	Min	SIM   
\# 	                         	Req (s) 	Description 	Mnemonic    	EGSE        	Key         	Description 	Mnemonic    	Value       	      	EGSE        	Verifier        	              	                         	            	Con	Fmt	Alt	Pos   
   	5.5\_V2.0                	    	            	            	            	            	            	            	            	      	            	            	            	                         	            	   	   	   	      
1.0	Turn on Pressure XDCRs 	    	            	            	            	            	            	            	            	                                           	            	            	            	                         	            	   	   	   	      
1.1	turn on Pressure Transducer A  	1.0   	XDCR A on   	1PRESAON    	            	            	Pressure XDCR A    	1PRDBAON    	ON          	      	            	            	           	                         	            	Y  	6            	   	      
1.2	turn on Pressure Transducer B  	1.0   	XDCR B on   	1PRESBON    	            	            	Pressure XDCR B   	1PRDBBON    	ON           	      	            	            	           	                         	            	Y  	6            	   	      
2.0	Open Valves              	    	            	            	            	            	            	            	            	      	            	            	                	                         	            	   	   	   	      
2.1	Enable valve A           	1.0  	Enable valve A	1LVCAEN     	            	            	Little valve Enabled A	1LVDBAON    	ENB           	      	            	            	            	                         	            	Y  	6  	   	      
2.2	Enable valve B           	1.0  	Enable valve B	1LVCBEN     	            	            	Little valve enabled B	1LVDBBON    	ENB           	      	            	            	            	                         	            	Y  	6          	   	      
2.3	Open valve A             	1.0   	Open valve  A	1LVCOAON    	            	            	Little valve open A	NONE        	            	      	            	            	           	                         	            	Y  	6           	   	      
2.4	Open valve B             	1.0   	Open valve  B	1LVCOBON    	            	            	Little valve open B	NONE        	            	      	            	            	          	                         	            	Y  	6          	   	      
3.0	Disable Valves           	    	            	            	            	            	            	            	            	      	            	            	            	                         	            	   	   	   	      
3.1	Disable valve A          	1.0   	Disable A   	1LVCADS     	            	            	Disable A   	1LVDBAON    	DIS           	      	            	            	            	                         	            	Y 	6          	   	      
3.2	Disable valve B          	1.0   	Disable B   	1LVCBDS     	            	            	Disable B   	1LVDBBON    	DIS          	      	            	            	           	                         	            	Y   	6          	   	      
4.0	Monitor Pressure as available           	    	            	            	            	            	            	            	            	      	            	            	            	                         	            	   	   	   	      
4.1	Measure Pressure A       	  	            	            	            	            	Differential Housing  	1HOPRAPR    	-12-- +2       	T     	            	            	 	If P<-10~T, OBF may be damaged,  	            	  	6         	   	      
   	                         	    	            	            	            	            	 Pressure A  	            	            	      	            	            	 	consult ACIS engineering team.                         	            	   	         	   	      
4.2	Measure Pressure B       	    	            	            	            	            	Differential Housing  	1HOPRBPR    	-12-- +2      	T     	            	            	  	Monitor pressure whenever data is   	            	  	6         	   	      
   	                         	    	            	            	            	            	Pressure B  	            	            	      	            	            	           	available.                          	            	   	         	   	      
 	Total Time	8	 	 	 	 	 	 	 	 	 	 	 	 	 	 	 	 	 