Step	Title                    	Time	Command     	Command     	Cmd         	Seq         	Telemetry   	Telemetry   	Expected    	Units 	Telemetry   	Other       	Crit        	Description              	Notes       	RT 	Tlm	Min	SIM   
\# 	                         	Req (m)	Description 	Mnemonic    	EGSE        	Key         	Description 	Mnemonic    	Value       	      	EGSE        	Verifier        	              	                         	            	Con	Fmt	Alt	Pos   
   	5.10\_V1.1               	    	            	            	            	            	            	            	            	      	            	            	            	                         	            	   	   	   	      
1.0	Set FP Temperature to +20 C           	    	            	            	            	            	            	            	            	      	            	            	            	                         	            	   	   	   	      
1.1	verify power consumption  	10   	            	            	            	            	DEA Input I 	1DEICACU    	0.60--0.78           	A      	            	            	2            	DEA A Power should be $21\pm3$~W, 	            	Y  	6  	   	      
   	                         	    	            	            	            	            	DEA Input V 	1DE28AVO    	27.0-35.0           	V      	            	            	           	current is noisy so an avg. is needed 	            	Y  	6            	   	      
1.2	Set warm boot flag off   	1   	DPA Cold Boot	1WRMBTSB(0) 	            	            	            	            	            	      	            	            	            	disable Warm boot if BEP watchdogs	   	Y 	6 	   	      
1.3	Upload bakeout patch     	1   	writeBep    	WBBOEN\_ENB 	            	            	 	            	            	      	Check cmdResult == OK 	            	A 	Enable the "enable"      	            	Y  	6 	   	      
   	                         	    	            	            	            	            	            	            	            	      	commandEcho 101 	            	            	  	  	   	   	   	      
1.4	Enable the FP BO htr     	1   	changeConfigSetting	WSBAKENABL  	            	            	  	            	            	      	Check cmdResult == OK 	            	A          	 	 	Y  	6  	   	      
   	                         	    	            	            	            	            	            	            	            	      	commandEcho 114 	            	            	  		   	   	   	      
1.5	Set FP temp to +20~C     	1   	changeConfigSetting	WSFTPOS\_20     	            	            		            	            	      	Check cmdResult == OK   	            	A         	Set desired FP temp to +20~C, 	 	Y  	6 	   	      
 	     	   	 	     	            	            		            	            	      	 commandEcho 225  	            	        	FP bakeout htr comes on. 	  	 	 	   	      
2.0	Monitor Cold Radiator Ts           	    	            	            	            	            	            	            	            	      	            	            	            	                         	            	   	   	   	      
2.1 	Check Cold Radiator Temps    	    	            	            	            	            	Cold Radiator T A 	 1CRAT           	 $\sim+10$  	C     	            	 	            	cold radiator temp should stabilize 	            	  	6        	   	      
   	                         	    	            	            	            	            	Cold Radiator T B 	 1CRBT           	 $\sim+10$  	C     	            	 	            	 	            	  	            	   	      
 	Total Time	14	 	 	 	 	 	 	 	 	 	 	 	 	 	 	 	 	 